HIGHLIGHTS
- What: All the claims by the vendor were successfully verified with and addition that vendor claims 12-hour stability while the study showed a stability of 18 hours. In precision the coefficient of variance in the experiment showed better results. All the experiments intercept and slope both are in allowable limits hence experiment was considered pass. - (time from draw) 0 hours, 1 hour, 4-hour, 8-hour, 12 hours, 24 hours). Similarly, the study showed that the samples are stable for a period of 13.2 hour at room temperature and can be analyzed even without mixing . . .

If you want to have access to all the content you need to log in!
Thanks :)
If you don't have an account, you can create one here.